Vulnerabilidades do Dräger Perseus A500
🔔 Criar alerta RSS Verificador de versão
CVEs publicadas que afetam Dräger Perseus A500: 1 registro(s), com versões afetadas e correções.
CVE-2019-25723
Média · 4.0
CWE-1286 02/06/2026
Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.