CVE-2025-14095 — Privilege boundary violation in Radiometer Products
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software. Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
Descrição oficial em inglês (tradução ainda não revisada).
Quem explora e qual a prioridade
Nota de 0 a 100 que combina exploração confirmada, uso por ransomware e grupos APT, exploits públicos, EPSS e severidade.
Como a nota foi calculada
- EPSS 0.2% (probabilidade de exploração em 30 dias)+1
- Severidade técnica (medium)+3
A nota ordena a fila de correção; não substitui o contexto dos seus ativos (exposição à internet, criticidade, compensações).
Triagem
Está sendo explorada, qual a probabilidade de exploração, o que afeta e qual a severidade segundo as fontes.
Avaliação SSVC da CISA CISA
Três pontos de decisão que a CISA publica para as CVEs que avalia (SSVC 2.0.3). Uma decisão de priorização, não uma nota de severidade.
Remediação
Correções e mitigações documentadas pelas fontes — com link para o aviso original.
Nenhuma correção documentada pelas fontes consultadas. Isso não significa que não exista: verifique o aviso do fabricante.
Primeiras 24 horas
Ordenado a partir dos próprios campos do registro — exposição primeiro, porque não se corrige o que não se encontrou. Ponto de partida, gerado por regras.
- Identifique ativos expostos que executam as combinações de fabricante/produto/versão afetadas (Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers, Radiometer Medical Aps ABL9 Analyzers, Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers…).
- Nenhuma correção documentada nas fontes: reduza a exposição e monitore o aviso do fabricante.
- Procure nos logs disponíveis sinais de tentativa de exploração compatíveis com a classe da falha (controle de acesso), erros e anomalias de autenticação.
Escopo afetado
Fabricante, produto e versões como as fontes os descrevem.
| Fabricante | Produto | Versões | Status | Fonte |
|---|---|---|---|---|
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX analyzers | versão Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlyinversão Application software versions < 6.20 MR2 with Windows 10 as underlying OS | Vulnerável | CVE Program (cve.org) e CISA ADP |
| Radiometer Medical Aps | ABL9 Analyzers | versão Application software versions < 1.5.0 | Vulnerável | CVE Program (cve.org) e CISA ADP |
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | versão All application software versions with Windows 7, Windows XP as underlying OSversão Application software versions < 3.5MR11 with Windows 10 as underlying OS | Vulnerável | CVE Program (cve.org) e CISA ADP |
| Radiometer Medical Aps | AQT90 FLEX Analyzers | versão All Application software versions <= 8.13 MR2 | Vulnerável | CVE Program (cve.org) e CISA ADP |
Características do ataque
O vetor CVSS decodificado. Descreve o ataque, não a sua exposição a ele.
Todas as pontuações coletadas
As fontes pontuam de forma independente e podem discordar; cada linha diz quem pontuou e em qual versão.
| Nota | Versão | Severidade | Explor. | Impacto | Pontuado por |
|---|---|---|---|---|---|
| 6.8 | CVSS 3.1 | Média | 0.9 | 5.9 | 46b595e9-1acc-41cb-9398-adaf98d37a9b via NIST National Vulnerability Database |
| 6.8 | CVSS 3.1 | Média | — | — | Radiometer via CVE Program (cve.org) e CISA ADP |
| 5.7 | CVSS 3.1 | Média | — | — | Radiometer via CVE Program (cve.org) e CISA ADP |
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (nota exibida no topo: 46b595e9-1acc-41cb-9398-adaf98d37a9b) Fraqueza MITRE
Classe de fraqueza segundo o catálogo CWE do MITRE.
EPSS ao longo do tempo FIRST
13/10/2026 → 14/10/2026 · 2 leitura(s) · máximo 0.17%
Detecção
Derivado do vetor CVSS e da classe da fraqueza. Pontos de partida, não regras testadas.
- Vetor local/físico: concentre-se em telemetria de endpoint (EDR) dos hosts afetados.
- Audite ações administrativas e acessos a recursos fora do padrão do usuário.
Linha do tempo
O que aconteceu com esta CVE, do mais recente ao mais antigo.
- 2026
- Produtos afetados adicionados · radiometer medical aps/abl90 flex and abl90 flex plus analyzers, radiometer medical aps/aqt90 flex analyzers, radiometer
- Avaliação SSVC atualizada
- Ficha publicada
- CVSS alterado · 6.8
- Status na fonte alterado · → Deferred
- Registro criado nesta base
- 2025
- CVE publicada
Atividade das fontes
Leituras que se repetem foram contadas em vez de listadas.
- 1× Descrição oficial atualizada · 13/10/2026 – 13/10/2026
Referências
1 no registro
Em outras páginas do site
CVEs relacionadas (mesmos produtos)
- Alta CVE-2025-14097 Remote Code Execution Vulnerability in Radiometer Products 17/12/2025
- Alta CVE-2025-14096 Credential Disclosure vulnerability in Radiometer Products 17/12/2025
Não consta em nenhuma fonte consultada
Listado em vez de deixado em branco: um campo vazio e um não medido parecem iguais na tela, e só um deles é motivo para procurar em outro lugar.
- Nenhuma versão corrigida, patch ou mitigação documentada.
- Nenhum IOC confirmado (IPs, domínios, hashes ou artefatos de malware) fornecido.
- Nenhum inventário de ativos, controle compensatório ou evidência de implantação de patch da sua organização.
- Nenhuma captura de tráfego de exploração, amostra de log ou caso de incidente.
Respondido a partir deste registro
Texto montado apenas com os campos acima — sem afirmações adicionais.
O que os defensores devem saber primeiro?
CVE-2025-14095 é "Privilege boundary violation in Radiometer Products", de severidade Média. Afeta ABL800 BASIC and ABL800 FLEX analyzers (Radiometer Medical Aps); as versões estão na tabela de escopo. Não está na CISA KEV, o que não exclui exploração. CVSS 6.8 descreve a severidade técnica. EPSS 0.17% estima a probabilidade de exploração em 30 dias; nenhum dos dois substitui o contexto dos seus ativos. Verifique primeiro se ativos expostos ou críticos executam as versões afetadas, aplique as correções ou mitigações do fabricante e preserve logs para análise.