CVE Intelligence
Medium

CVE-2019-25717 — Dräger Infinity Delta/Kappa Patient Monitors Unauthenticated Log File Disclosure

Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.

Published Jun 2, 2026 · Updated Oct 13, 2026 · Sources: NIST National Vulnerability Database, CVE Program (cve.org) e CISA ADP · CNA: VulnCheck

Who exploits it and how urgent it is

A 0–100 score combining confirmed exploitation, ransomware and APT use, public exploits, EPSS and severity.

No exploitation signal in the sources consulted so far.
How the score was calculated
  • EPSS 0.2% (30-day exploitation probability)+1
  • Technical severity (medium)+3

The score orders the patch queue; it does not replace your asset context (internet exposure, criticality, compensating controls).

Triage

Is it exploited, how likely is exploitation, what does it touch, and how severe do the sources call it.

Exploitation
Not confirmed
Not in CISA KEV — an absence of record, not proof it is unexploited.
EPSS FIRST
0.2%
ahead of 10% of scored CVEs · Oct 14, 2026
Affects
Draeger
Delta Xl Firmware and 5 more
CVSS base
4.3
Medium · 2 source(s)

CISA SSVC assessment CISA

Three decision points CISA publishes for the CVEs it assesses (SSVC 2.0.3). A stakeholder decision, not a severity score.

Exploitation
None
none · proof-of-concept · active
Automatable
No
can an attacker script the kill-chain steps
Technical impact
Partial
partial · total control of the vulnerable component

Remediation

Fixes and mitigations documented by the sources — linked to the original advisory.

No fix documented by the sources we poll. That does not mean none exists: check the vendor advisory.

First 24 hours

Ordered from the record’s own fields — exposure first, because you cannot patch what you have not found. A rule-generated starting point.

  1. Identify exposed assets running the affected vendor/product/version combinations (Draeger Delta Xl Firmware, Draeger Infinity Delta Firmware, Draeger Kappa Firmware…).
  2. No fix documented in the sources: reduce exposure and monitor the vendor advisory.
  3. Search available logs for exploit probes, errors and authentication anomalies.

Affected scope

Vendor, product and versions as the sources word them.

VendorProductVersionsStatusSource
Draeger Delta Xl Firmware versions not specified by the source Vulnerable NIST National Vulnerability Database
Draeger Infinity Delta Firmware versions not specified by the source Vulnerable NIST National Vulnerability Database
Draeger Kappa Firmware versions not specified by the source Vulnerable NIST National Vulnerability Database
Dräger Infinity Delta version all software versions Vulnerable CVE Program (cve.org) e CISA ADP
Dräger Infinity Delta XL version all software versions Vulnerable CVE Program (cve.org) e CISA ADP
Dräger Infinity Kappa version all software versions Vulnerable CVE Program (cve.org) e CISA ADP

Attack characteristics

The CVSS vector, decoded. It describes the attack, not your exposure to it.

Attack vector
Adjacent
Attack complexity
Low
Privileges required
None
User interaction
None
Scope
Unchanged
Confidentiality
Low
Integrity
None
Availability
None

Every base score collected

Sources score independently and may disagree; each row says who scored it and under which version.

ScoreVersionSeverityExpl.ImpactScored by
5.3 CVSS 4.0 Medium — — disclosure@vulncheck.com via NIST National Vulnerability Database
5.3 CVSS 4.0 Medium — — VulnCheck via CVE Program (cve.org) e CISA ADP
4.3 CVSS 3.1 Medium 2.8 1.4 disclosure@vulncheck.com via NIST National Vulnerability Database
4.3 CVSS 3.1 Medium — — VulnCheck via CVE Program (cve.org) e CISA ADP
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N (score shown at the top: disclosure@vulncheck.com)

Weakness MITRE

Weakness class according to MITRE’s CWE catalog.

CWE-538

Other CVEs with this weakness →

EPSS over time FIRST

Oct 13, 2026 → Oct 14, 2026 · 2 reading(s) · max 0.20%

Detection

Read off the CVSS vector and the weakness class. Starting points, not tested rules.

  • Not enough vector or weakness data: use the affected application’s logs as a starting point.
application logs

Timeline

What happened to this CVE, newest first.

  • 2026
  • Affected products added · dräger/infinity delta, dräger/infinity delta xl, dräger/infinity kappa
    Oct 13, 2026 · CVE Program (cve.org) e CISA ADP
  • SSVC assessment updated
    Oct 13, 2026 · CVE Program (cve.org) e CISA ADP
  • Record published
    Oct 13, 2026
  • CVSS changed · 4.3
    Oct 13, 2026 · NIST National Vulnerability Database
  • Source status changed · → Analyzed
    Oct 13, 2026 · NIST National Vulnerability Database
  • Record created in this database
    Oct 13, 2026 · NIST National Vulnerability Database
  • CVE published
    Jun 2, 2026 · VulnCheck

Source activity

Repeated readings are counted rather than listed.

  • 1× Official description updated · Oct 13, 2026 – Oct 13, 2026

References

2 on the record

Elsewhere on this site

Draeger
every CVE for this vendor
CWE-538
other pages naming this weakness

Related CVEs (same products)

  • High CVE-2019-25716 Dräger Infinity Delta/Kappa Patient Monitor DoS via Malformed Network Packet Jun 1, 2026

Not in any source we poll

Listed rather than left blank: an empty field and an unmeasured one look identical on screen, and only one is a reason to look elsewhere.

  • No fixed version, patch or mitigation documented.
  • No confirmed IOCs, IP addresses, domains, file hashes, or malware artifacts supplied.
  • No organization-specific asset inventory, compensating-control status, or patch deployment evidence supplied.
  • No exploit packet captures, log samples, or incident case IDs supplied.

Answered from this record

Text assembled only from the fields above — no additional claims.

What should defenders know first?

CVE-2019-25717 is "Dräger Infinity Delta/Kappa Patient Monitors Unauthenticated Log File Disclosure", rated Medium. It affects Delta Xl Firmware (Draeger); versions are in the scope table. It is not in CISA KEV, which does not rule out exploitation. CVSS 4.3 describes technical severity. EPSS 0.20% estimates the probability of exploitation in 30 days; neither replaces your asset context. First verify whether exposed or business-critical assets run the affected versions, apply vendor fixes or mitigations, and preserve logs for review.

Comments

Moderated technical discussion. Read the community guidelines.

No comments yet.

Sign in or create an account to comment and suggest corrections.