CVE-2019-25723
Medium · 4.0
CWE-1286 Jun 2, 2026
Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.