Thermofisher Ion Torrent Onetouch 2 Firmware vulnerabilities
🔔 Create alert RSS Version checker
Published CVEs affecting Thermofisher Ion Torrent Onetouch 2 Firmware: 2 record(s), with affected versions and fixes.cpe:2.3:o:thermofisher:ion_torrent_onetouch_2_firmware
CVE-2025-54304
Critical · 9.8
CWE-200 Dec 4, 2025
Exposure of Sensitive Information in Thermofisher Ion Torrent Onetouch 2 Firmware
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
CVE-2025-53963
Critical · 9.8
CWE-521 Dec 4, 2025
Weak Password Requirements in Thermofisher Ion Torrent Onetouch 2 Firmware
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.