CVE Intelligence

Mirion vulnerabilities

🔔 Create alert RSS

Published CVEs affecting Mirion products: 5 record(s), with severity, exploitation and fixes.

Exploited only (KEV) →

CVE-2025-64778 High · 7.8 CWE-798 Dec 2, 2025
Mirion Medical EC2 Software NMIS BioDose Use of Hard-coded Credentials
NMIS/BioDose software V22.02 and previous versions contain executable binaries with plain text hard-coded passwords. These hard-coded passwords could allow unauthorized access to both the application and database.
CVE-2025-64642 High · 7.8 CWE-732 Dec 2, 2025
Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource
NMIS/BioDose V22.02 and previous versions' installation directory paths by default have insecure file permissions, which in certain deployment scenarios can enable users on client workstations to modify the program executables and libraries.
CVE-2025-64298 High · 7.5 CWE-732 Dec 2, 2025
Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource
NMIS/BioDose V22.02 and previous version installations where the embedded Microsoft SQLServer Express is used are exposed in the Windows share accessed by clients in networked installs. By default, this directory has insecure directory paths that allow access to the SQL Server database and configuration files, which can contain sensitive data.
CVE-2025-62575 High · 8.8 CWE-732 Dec 2, 2025
Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource
NMIS/BioDose V22.02 and previous versions rely on a Microsoft SQL Server database. The SQL user account 'nmdbuser' and other created accounts by default have the sysadmin role. This can lead to remote code execution through the use of certain built-in stored procedures.
CVE-2025-61940 High · 8.8 CWE-603 Dec 2, 2025
Mirion Medical EC2 Software NMIS BioDose Use of Client-Side Authentication
NMIS/BioDose V22.02 and previous versions rely on a common SQL Server user account to access data in the database. User access in the client application is restricted by a password authentication check in the client software but the underlying database connection always has access. The latest version of NMIS/BioDose introduces an option to use Windows user authentication with the database, which would restrict this database connection.