CVE Intelligence

Spacelabs Healthcare vulnerabilities

🔔 Create alert RSS

Published CVEs affecting Spacelabs Healthcare products: 1 record(s), with severity, exploitation and fixes.

Exploited only (KEV) →

CVE-2026-0611 Critical · 9.8 CWE-306 Jun 2, 2026
Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.