CVE Intelligence
High Authentication

CVE-2025-14097 — Remote Code Execution Vulnerability in Radiometer Products

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Published Dec 17, 2025 · Updated Oct 13, 2026 · Sources: NIST National Vulnerability Database, CVE Program (cve.org) e CISA ADP · CNA: Radiometer

Who exploits it and how urgent it is

A 0–100 score combining confirmed exploitation, ransomware and APT use, public exploits, EPSS and severity.

No exploitation signal in the sources consulted so far.
How the score was calculated
  • EPSS 0.5% (30-day exploitation probability)+6
  • Technical severity (high)+6

The score orders the patch queue; it does not replace your asset context (internet exposure, criticality, compensating controls).

Triage

Is it exploited, how likely is exploitation, what does it touch, and how severe do the sources call it.

Exploitation
Not confirmed
Not in CISA KEV — an absence of record, not proof it is unexploited.
EPSS FIRST
0.47%
ahead of 39% of scored CVEs · Oct 14, 2026
Affects
Radiometer Medical Aps
ABL800 BASIC and ABL800 FLEX analyzers and 2 more
CVSS base
7.2
High · 2 source(s)

CISA SSVC assessment CISA

Three decision points CISA publishes for the CVEs it assesses (SSVC 2.0.3). A stakeholder decision, not a severity score.

Exploitation
None
none · proof-of-concept · active
Automatable
No
can an attacker script the kill-chain steps
Technical impact
Total
partial · total control of the vulnerable component

Remediation

Fixes and mitigations documented by the sources — linked to the original advisory.

No fix documented by the sources we poll. That does not mean none exists: check the vendor advisory.

First 24 hours

Ordered from the record’s own fields — exposure first, because you cannot patch what you have not found. A rule-generated starting point.

  1. Identify exposed assets running the affected vendor/product/version combinations (Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers, Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers, Radiometer Medical Aps AQT90 FLEX Analyzers).
  2. No fix documented in the sources: reduce exposure and monitor the vendor advisory.
  3. Search available logs for exploit probes matching the vulnerability class (authentication), errors and authentication anomalies.

Affected scope

Vendor, product and versions as the sources word them.

VendorProductVersionsStatusSource
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers version Application software versions < 6.20MR2 with Windows 7, Windows XP operating sys Vulnerable CVE Program (cve.org) e CISA ADP
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers version Application software versions < 3.5MR11 with Windows 7, Windows XP operating sys Vulnerable CVE Program (cve.org) e CISA ADP
Radiometer Medical Aps AQT90 FLEX Analyzers version Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating s Vulnerable CVE Program (cve.org) e CISA ADP

Attack characteristics

The CVSS vector, decoded. It describes the attack, not your exposure to it.

Attack vector
Network
Attack complexity
Low
Privileges required
High
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

Every base score collected

Sources score independently and may disagree; each row says who scored it and under which version.

ScoreVersionSeverityExpl.ImpactScored by
7.2 CVSS 3.1 High 1.2 5.9 46b595e9-1acc-41cb-9398-adaf98d37a9b via NIST National Vulnerability Database
7.2 CVSS 3.1 High — — Radiometer via CVE Program (cve.org) e CISA ADP
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H (score shown at the top: 46b595e9-1acc-41cb-9398-adaf98d37a9b)

Weakness MITRE

Weakness class according to MITRE’s CWE catalog.

CWE-287 · Improper Authentication

Other CVEs with this weakness →

EPSS over time FIRST

Oct 13, 2026 → Oct 14, 2026 · 2 reading(s) · max 0.47%

Detection

Read off the CVSS vector and the weakness class. Starting points, not tested rules.

  • Prioritize edge telemetry for the network-reachable component (proxy, WAF, IDS/IPS).
  • Review authentication logs: logins without MFA, anomalous sessions and recently created accounts.
web/proxy/WAF logsnetwork IDS/IPSauthentication logs

Timeline

What happened to this CVE, newest first.

  • 2026
  • Affected products added · radiometer medical aps/abl90 flex and abl90 flex plus analyzers, radiometer medical aps/abl800 basic and abl800 flex ana
    Oct 13, 2026 · CVE Program (cve.org) e CISA ADP
  • SSVC assessment updated
    Oct 13, 2026 · CVE Program (cve.org) e CISA ADP
  • Record published
    Oct 13, 2026
  • CVSS changed · 7.2
    Oct 13, 2026 · NIST National Vulnerability Database
  • Source status changed · → Deferred
    Oct 13, 2026 · NIST National Vulnerability Database
  • Record created in this database
    Oct 13, 2026 · NIST National Vulnerability Database
  • 2025
  • CVE published
    Dec 17, 2025 · Radiometer

Source activity

Repeated readings are counted rather than listed.

  • 1× Official description updated · Oct 13, 2026 – Oct 13, 2026

References

1 on the record

Elsewhere on this site

Radiometer Medical Aps
every CVE for this vendor
Authentication
same class
CWE-287
other pages naming this weakness

Related CVEs (same products)

  • High CVE-2025-14096 Credential Disclosure vulnerability in Radiometer Products Dec 17, 2025
  • Medium CVE-2025-14095 Privilege boundary violation in Radiometer Products Dec 17, 2025

Not in any source we poll

Listed rather than left blank: an empty field and an unmeasured one look identical on screen, and only one is a reason to look elsewhere.

  • No fixed version, patch or mitigation documented.
  • No confirmed IOCs, IP addresses, domains, file hashes, or malware artifacts supplied.
  • No organization-specific asset inventory, compensating-control status, or patch deployment evidence supplied.
  • No exploit packet captures, log samples, or incident case IDs supplied.

Answered from this record

Text assembled only from the fields above — no additional claims.

What should defenders know first?

CVE-2025-14097 is "Remote Code Execution Vulnerability in Radiometer Products", rated High. It affects ABL800 BASIC and ABL800 FLEX analyzers (Radiometer Medical Aps); versions are in the scope table. It is not in CISA KEV, which does not rule out exploitation. CVSS 7.2 describes technical severity. EPSS 0.47% estimates the probability of exploitation in 30 days; neither replaces your asset context. First verify whether exposed or business-critical assets run the affected versions, apply vendor fixes or mitigations, and preserve logs for review.

Comments

Moderated technical discussion. Read the community guidelines.

No comments yet.

Sign in or create an account to comment and suggest corrections.